4
Enter the channel ID.
Press the Channel key, and enter the channel ID.
Use the numeric keypad to enter the 4-digit medical telemetry channel
ID. The set channel ID will be displayed on the upper left of the home
display.
5
Enter the group ID.
Press the Group key, and enter the group ID.
Use the numeric keypad to enter the group ID in the range of 00 to 63.
6
Save the channel ID and group ID.
Pressing the Save key will store the channel ID and group ID.
Verify that the “Complete” message is displayed.
If an error is found on the password, channel ID, or group ID, the following message will be displayed.
“Invalid password”
: The entered password is incorrect. Enter the password again and press
the Save key.
“Invalid data”
: The entered channel ID or group ID is outside the allowable range.
Enter the ID again and press the Save key.
7
Verify the stored channel ID and group ID.
8
Press the Telemetry Wave key to display the Telemetry Wave Setup menu.
Select the transmitting waveform from ECG1 or
ECG1+2 .
ECG1 will transmit ECG1, RESP, CO
2
, BP1, BP2, and SpO
2
.
However, RESP waveform will not be transmitted if APNEA
source is CO
2
.
ECG1+2 will transmit ECG1, ECG2, RESP/CO
2
, BP1, and
SpO
2
.
One of the waveform from CO
2
and RESP will be transmitted
depending on the APNEA source selection.
8
-
16
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...