FAST SAT Setup
By selecting ON for “FAST SAT”, abrupt change of the SpO
2
value can be monitored.
N O T E
To pick up the abrupt change of the value sooner, and to take advantage of the
qualities of FAST SAT mode, it is recommended to set 2-4sec for SpO
2
averaging time when FAST SAT is set to ON.
1
Press the Setup
→
Page Down key.
The setup menu to set the FAST SAT will be displayed.
2
Select from ON or OFF .
●
PI Display
Whether or not to display the PI (Perfusion Index) data can be selected.
The perfusion index is calculated by the following formula; pulsatile signal
÷
apulsatile signal
×
100,
and indicates the patient’s circulation condition. This can be used to find a good perfusion site to
attach the sensor. Also, it can be used as diagnosis index to predict the patient’s critical condition
when at low perfusion.
1
Press the Setup
→
Page Down key.
The setup menu to set the PI display will be displayed.