Dyna Alert Function
Using a cuff allows to measure the blood pressure noninvasively, but on the other hand, there is a
demerit of not being able to perform the measurement continuously. Therefore, there is always a risk
of sudden blood pressure change in between the periodic measurements.
The Dyna Alert function is a technology to prevent accidents which may occur by this sudden BP
change during the non-measured duration by estimating the variation of circulatory dynamics using the
parameters obtained from ECG and photoplethysmogram (SpO
2
), and initiating a new NIBP
measurement if a change in the circulatory dynamics is detected.
Parameters used for Dyna Alert Function
DT
RR
a
b d
ECG
PTG
Second
Derivative
PTG
The Dyna Alert function will operate on the following measurement condition.
Patient Classification
: Adult (20kg or above)
Cuff Applied Site
: Upper arm
SpO
2
Sensor Applied Site
: Fingertip
NIBP Measurement Interval : 5 to 60 minutes
C A U T I O N
When a PTG (SpO
2
) sensor is applied to the toe or forehead, the Dyna Alert
may not function depending on the patient’s condition.
1
Select ON/OFF for Dyna Alert function.
ON will set the Dyna Alert function ON.
OFF will set the Dyna Alert function OFF.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...