7
-
10
Trend Display
In this section, procedure to operate and print the trend display is explained.
The data of 24 hours will be automatically stored and displayed in 1 minute interval for the parameters
displayed on the home display.
1
Press the Table/Trend key to display the trend data.
The trend display is composed of 2 pages which can be switched by using the Page Down or
Page Up keys.
The graph is divided into upper and lower row on each page.
On each row, a graph trend of 4 parameters can be displayed simultaneously.
2
Set the display interval for the trend data.
Pressing the Interval xH key will display the selection tool to select the display
interval.
Select from 1H / 2H / 4H / 8H / 12H / 24H .
Time Range
Sample Rate
1 hour
1 minute
2 hours
2 minutes
4 hours
4 minutes
8 hours
8 minutes
12 hours
12 minutes
24 hours
24 minutes
Scale
Real Time Scale
Current stored data amount (filled-in part) of the 24-hour data.
Scale
Switches the page.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...