6
-
53
3
If 60M or 120M is selected, ON/OFF of “Start 5min. early” can be selected.
Press the Setup
→
Page Down keys to display the second page of the NIBP setup menu.
If 60M or 120M is selected for interval, the
measurement can be started 5 minutes early.
If outputting the data to PC or other external equipment using
the PC communication function of this device, an error may be
generated to the NIBP measurement time depending on the
input interval of the external equipment. As this device outputs
the data at completion of NIBP measurement, if the external
equipment inputs the data at 60 minutes interval, 60 minutes
time lag will occur. By starting the measurement 5 minutes
earlier, this time lag between the external equipment can be
minimized.
NIBP 1-Minute Interval Measurement
The 1-minute interval measurement will automatically stop after 12 minutes and returns to the
automatic measurement mode of 2.5 minutes interval.
1
Press the
(NIBP Auto Mode) key.
The measurement interval selection for the automatic
measurement will be displayed.
2
Selecting 1M will immediately start the 1-minute interval measurement.
To suspend the measurement, press the
(NIBP Start/Stop) key.
N O T E
Pressing the
(NIBP Start/Stop) key will only suspend the 1-minute
interval measurement. To cancel the 1-minute interval measurement
operation, change the measurement interval.
6
Non
-In
va
sive Blood Pre
ssu
re
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...