11
-
34
Item
Default
Backup
HR
ON, 60 to OFF
ASYSTOLE ON,
5sec.
VF ON
VT ON
SLOW VT
OFF
RUN OFF,
3beats
COUPLET OFF
PAUSE OFF,
3.0sec.
BIGEMINY OFF
TRIGEMINY OFF
FREQUENT OFF,
10beats/min.
TACHY OFF
BRADY OFF
ST
Ι
,
ΙΙ
,
ΙΙΙ
, aVR, aVL,
aVF, V
All Alarms OFF
Indiv. Alarm OFF
OFF to OFF
BP1
ON
S
80 to 180mmHg/10 to 27kPa
D
OFF to OFF
M
OFF to OFF
BP2 to BP6
OFF
S
OFF to OFF
D
OFF to OFF
M
OFF to OFF
RR
ON, 6 to 30
APNEA ON,
60sec.
SpO
2
ON, 90 to OFF
SEC Alarm
(Nellcor Model:
DS-7000)
OFF
NIBP
ON
S
70 to OFF
D
OFF to OFF
M
OFF to OFF
TEMP1 to TEMP3,
Tb
OFF
OFF to OFF
EtCO
2
OFF
OFF to OFF
In-CO
2
OFF
OFF
Ex-O
2
OFF
OFF to OFF
FiO
2
OFF
18 to OFF
Ex-N
2
O
OFF
OFF to OFF
In-N
2
O
OFF
OFF to OFF
Ex-AGT
OFF
OFF to OFF
In-AGT
OFF
OFF to OFF
Alarm Mode
1 to 5 (Child)
MAC OFF
OFF
{