6
-
69
Vigilance Data
When Vigilance, Vigilance CEDV, Vigilance
ΙΙ
, or Vigileo (oximeter/CCO measurement device
manufactured by Baxter) is used, Vigilance data display can be selected from several modes.
【
ICO Mode
】
STAT Mode : When Vigilance is in CCO mode, STAT mode display can be set ON or OFF.
Index Disp. : When Vigilance is in CCO mode, Index display can be set ON or OFF.
When the Vigilance is in ICO mode, the 6 latest data of ICO (Intermittent Cardiac Output) and ICI
(Intermittent Cardiac Index) will be displayed.
STAT Mode / Index Display
1
ON / OFF for STAT Mode and ON / OFF for Index display can be selected on the
Vigilance display.
【
STAT Mode OFF , Index Display OFF
】
SvO
2
(or ScvO
2
), CCO, EDV, BT data will be displayed inside
the SvO
2
+CO numeric data box.
【
STAT Mode OFF , Index Display ON
】
By setting the Index display ON, CCI, EDVI data will be
displayed instead of CCO and EDV.
【
STAT Mode ON , Index Display OFF
】
By setting the STAT mode ON, CCO-STAT, EDV-STAT data
will be displayed instead of CCO and EDV.
【
STAT Mode ON , Index Display ON
】
By setting the STAT mode and Index display ON, CCI-STAT,
EDVI-STAT will be displayed instead of CCO and EDV.