Source Setup
On the “Source Setup” screen, whether or not to automatically switch the PR source, and the priority
of HR/PR alarm source can be set.
1
Press the Menu
→
Initial Settings
→
Source Setup keys.
The “Source Setup" screen will be displayed.
2
Select ON/OFF for “PR Source Auto Switch”.
When PR is selected for “Sync. Mark/Tone”, “Alarm
Source”, whether or not to automatically switch the PR
source can be set.
ON will automatically switch the PR source to
measurable parameter.
OFF will not automatically switch the PR source. The
PR source will be fixed according to the priority set on
the “PR Source Setup”.
C A U T I O N
In case of DS-LANII network, if the “PR Source” is BP, the ECG waveform will
not be transmitted. On the central monitor, PR-IBP value will be displayed for
HR. However, the HR value from ECG will be displayed on the NIBP list and ST
measurement table.
3
Set the “HR/PR Auto Alarm Source Priority”.
When the HR/PR alarm source is set to Auto , the
priority of the alarm source parameter can be selected.
The parameters inside the key are displayed in the
order of priority which is high from the left. Select the
priority order from the 6 selections.
For example, if [ECG/SpO
2
/BP] is selected, the HR/PR
source will be set in the priority of HR, PR from SpO
2
waveform, and PR from BP waveform.
N O T E
When the HR/PR alarm source is set to Auto , HR may not be displayed inside the
HR/PR numeric data box depending on the "HR/PR Auto Alarm Source Priority”
selection.
9
-
36
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...