6
-
77
●
Anesthetic Gas Alarm
In-AGT Alarm
The upper/lower limit can be set in 0.1% increments.
Lower
Limit
The adjustable range differs depending on the
anesthetic gas label.
ISO, HAL, ENF: 0.5 to 4.0%
DES: 0.5 to 12.0%
SEV: 0.5 to 5.0%
Turns OFF if the value below the adjustable range is
set.
Upper
Limit
The adjustable range differs depending on the
anesthetic gas label.
ISO, HAL, ENF: 3.0 to 6.0%
DES: 8.0 to 20.0%
SEV: 3.0 to 8.0%
Turns OFF if the value above the adjustable range is
set.
Auto
Automatically sets the upper limit to +0.8%, and lower
limit to –0.8% from the current value.
Ex-AGT Alarm
The upper/lower limit can be set in 0.1% increments.
Lower
Limit
The adjustable range differs depending on the
anesthetic gas label.
ISO, HAL, ENF: 0.5 to 4.0%
DES: 0.5 to 12.0%
SEV: 0.5 to 5.0%
Turns OFF if the value below the adjustable range is
set.
Upper
Limit
The adjustable range differs depending on the
anesthetic gas label.
ISO, HAL, ENF: 3.0 to 6.0%
DES: 8.0 to 20.0%
SEV: 3.0 to 8.0%
Turns OFF if the value above the adjustable range is
set.
Auto
Automatically sets the upper limit to +0.8%, and lower
limit to –0.8% from the current value.
●
MAC Value Alarm
The upper limit can be set in 0.1 increments.
Lower
Limit
(No setting)
Upper
Limit
Range: 0.1 to 9.9.
Turns OFF if 9.9 or above is set.
6
Multiga
s Data
(MGU-7
01/7
02/Poet IQ, MGU-80
1P/8
02/803
)
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...