C A U T I O N
y
For MGU-701/MGU-702/Poet IQ, do not sterilize the airway adapter using
autoclave methods.
y
For MGU-701/MGU-702/Poet IQ, do not reuse / re-sterilize the disposable
airway adapter.
y
For MGU-701/MGU-702/Poet IQ, water traps are single use only. Do not
attempt to drain the water trap when full.
y
For MGU-801P/MGU-802/MGU-803, do not use other than specified
cleaning methods for the water trap.
y
For MGU-801P/MGU-802/MGU-803, do not clean or wash the filter
housing of the water trap.
y
For MGU-801P/MGU-802/MGU-803, never allow alcohol to enter the filter
housing.
y
For MGU-801P/MGU-802/MGU-803, never force air through the water
trap.
y
Do not open the housing.
y
Do not allow liquids such as alcohol or cleaning solution enter the
equipment or connectors.
y
If you accidentally wet the device, dry it completely and verify it operates
safely before usage.
y
Replace the components periodically as specified.