6
-
68
Tb (Blood Temperature)
When thermodilution catheter is used to measure the cardiac output, the blood temperature can be
monitored.
Tb Alarm
: Sets ON/OFF of the blood temperature alarm and upper and lower alarm limits.
Tb Alarm
1
Select ON/OFF for blood temperature alarm and set upper and lower alarm limits.
The upper and lower limit can be set in increments of
0.5
°
C/1.0
°
F.
Key
Item
Description
Individual Alarm
Selecting ON will generate a TEMP alarm.
Selecting OFF will not generate a TEMP alarm.
▲
Current Value
Indicates the current value.
Lower Alarm Limit
Sets the lower alarm limit (30.0 to 44.0
°
C/86.0 to
111.0
°
F). Setting the value 30.0
°
C/86.0
°
F or below will
turn the alarm OFF.
Upper Alarm Limit
Sets the upper alarm limit (31.0 to 45.0
°
C/88.0 to
113.0
°
F). Setting the value 45.0
°
C/113.0
°
F or above will
turn the alarm OFF.
Automatic Setup
Automatically sets the upper limit to +2.0
°
C/+4.0
°
F from
the current value, and lower limit to –2.0
°
C/–4.0
°
F from
the current value.
To maintain the alarm setting even after the power is turned OFF or after a discharge procedure, store
the setting to one of the alarm modes, or select ON for “Backup at Discharge” on the Discharge
Mode setup menu under the Initial Settings.
Reference
For the alarm mode setup procedure, refer to “9. Initial Settings –Alarm Mode Setup–”.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...