Precautions about the Wired Network System (DS-LAN II/DS-LAN III)
W A R N I N G
z
Do not connect unspecified device to a wired network.
z
Do not mix devices with DS-LANII and DS-LANIII setting in the same
wired network. The network may cease and proper monitoring may not be
possible.
z
Before setting the bed ID, make sure that the DS-LAN
(DS-LANII/DS-LANIII) is correctly set on the Initial Settings menu. If not
correctly set, the network may cease which may lead to accidents such as
not transmitting life threatening alarms to the central monitor.
C A U T I O N
z
If performing wired network transmission, configure the display so that the
numeric data corresponded to the waveform displayed. If not, the
displayed waveform or numeric data may not be transmitted.
z
The Operation Room ID is factory set to 000. If connected to the wired
network with the ID unchanged, monitoring on the central monitor will not
be possible.
z
When connected to the wired network, verify that the Operation Room ID
is not duplicated with other monitors. Otherwise, monitoring on the central
monitor for both monitors will not be possible.
z
Make sure to set the Monitor ID in the following range.
y
For DS-LAN
ΙΙ
network: 001 to 048
y
For DS-LAN
ΙΙΙ
network: 001 to 100
z
Precautions about the DS-LANII Network System
y
Arrhythmia alarm of TACHY, BRADY, COUPLET, PAUSE,
TRIGEMINY will not be transmitted.
y
Arrhythmia alarm of “SLOW VT” will be transmitted as “VT”.
y
On a wired network, waveform, numeric data, alarm of TEMP3 will not
be transmitted. Also, the displayable waveform, numeric data, alarm
differs depending on the connected central monitor. Refer to the
operation manual for the respective central monitor.
y
If “HR/PR Alarm Source” is PR, and “PR Source” is BP, ECG waveform
will not be transmitted on a wired network. On the central monitor,
PR-IBP value will be displayed for HR. However, HR value from ECG
will be displayed for the NIBP list and ST measurement table.
y
If the “RR Alarm Source” selection is other than Impedance ,
respiration waveform will not be transmitted on a wired network.
y
If the “RR Alarm Source” selection is other than GAS , CO
2
waveform
will not be transmitted on a wired network.
y
For numeric data displayed as “
×××
”, maximum or minimum value of
measurable range will be transmitted.
y
The numeric data displayed as “
−
−
−
” will be treated as not measured
data.
xxvi
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...