Power Source and Ground Connection
Connect the power supply cable (CS-18) provided to the power supply connector on the DS-7000.
Connect the other end of the power cable to the hospital grade 3-way outlet with ground terminal.
Power Supply Connector
W A R N I N G
z
Use only the supplied 3-way AC power cable. Use of other cables may
result in electric shock to the patient and the operator.
z
The power cable must be connected to the hospital grade outlet.
z
When using multiple ME equipment simultaneously, perform equipotential
grounding to prevent potential difference between the equipment. Even a
small potential difference may result in electric shock to the patient and the
operator.
8
Powe
r Source and G
rou
nd
Conn
ectio
n
N O T E
Equipotential Grounding
When connecting multiple devices, electrical potential difference may be
generated between the devices. This may result in electric shock to the patient
connected to these devices. Pay special attention for use in the operating
room, ICU, CCU, Cardiac Catheter Laboratory, and Cardiovascular X-ray
room. To avoid such electrical potential difference, use the accessory ground
cable to connect each device’s potential equalization terminal to the same
ground terminal. This is called equipotential grounding.
8
-
3
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...