Wireless Network Construction
This section explains the procedure on how to use this equipment with a telemetry system.
By connecting the telemetry transmitter module (HLX-501), the DS-7000 can be connected to a
wireless network system and display the data on the central monitor.
DS-7600 Series Central Monitor
DS-7000 Bedside Monitor
D A N G E R
When monitoring a patient with wireless telemetry, make sure the patient data
is properly received at the central monitor. Pay special attention when the
channel ID at the bedside monitor is changed.
W A R N I N G
z
Some wireless combinations of telemetry transmitters may generate
interference with other devices.
z
Before selecting a channel, verify that it will not interfere with other
channels.
z
Make sure the telemetry manager of your system is aware of any changes
to the telemetry channels.
z
If transmitters are used in a neighboring medical facility, your facility and
the neighboring facility must make agreements on the setting of the
telemetry channels to prevent telemetry interference.
C A U T I O N
The setup of channel ID and group ID should be performed only by our
service representative. Users should not perform this procedure as
malfunction to the equipment may occur.