Zero Balance of Blood Pressure
1
Press the Menu
→
Function
→
BP Zero keys to display the zero balance screen.
2
Open the three-way valve of the pressure transducers to air.
When performing zero balance of all the pressure lines, open the three-way valve of all the
pressure transducers.
3
To perform zero balance for each pressure line, press the BP Zero key for the respective BP
for more than 3 seconds.
To perform zero balance for all pressure lines, press the Zero All BP key for more than 3
seconds.
<During zero balance>
<When zero balance is complete>
Verify that the BP waveform is positioned at zero, and “0” is
displayed for the BP value. A message, “Zero complete” will
be displayed when the procedure is complete.