11
-
13
Operating Temperature
: 10 to 35
°
C (with O
2
sensor), 10 to 40
°
C (without O
2
sensor)
Operating Humidity
: 10 to 95% RH non-condensing
Operating Atmospheric Pressure : 700 to 1200hPa (525 to 900mmHg) 700hPa = 3048m
Cardiac Output
Measurement Method : Thermodilution Method
Measurement Range : 0.1 to 20L/min
Measurement Accuracy : 0.1 to 10L/min:
less than
±
10%
10 to 20L/min:
less than
±
15%
Measurement Temperature Range and Accuracy
Blood Temperature
: 17 to 45
°
C
±
0.3
°
C
Injectate Temperature : –1 to 35
°
C
±
0.5
°
C
Recording
Recording Speed
: 50mm/s, 25mm/s (Error: within
±
5%)
Resolution
Head Direction : 8 dots/mm
Feed Direction : 40 lines/mm (at recording speed of 25mm/s)
Rec. Waveform
: 3 waveforms
Rec. Type
: Waveform Recording, Table Recording, Graphic Recording
Detection
: Paper out, page mark, paper cassette error, printhead temperature
Protective Circuit
: Printhead overcurrent, printhead overheating, motor overcurrent, surge
current
Analog Waveform Output
Output Voltage
: ECG output 1V/mV (fixed), BP output 1V/100mmHg (fixed)
Output Voltage Accuracy : within
±
10% (Both ECG and BP output)
Delay Time
: 35ms and below (Both ECG and BP waveform)
Frequency Range
: ECG output 0.05Hz to 30Hz
BP output DC to 30Hz
Pacemaker Pulse
: No Pacemaker Pulse
11
Specifi
cation /
Performan
ce
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...