C A U T I O N
z
The OR ID is factory set to “000”. If connected to a wired network with
the bed ID unchanged, monitoring on the central monitor will not be
possible.
z
When connecting to the DS-LAN
ΙΙ
network, make sure that there are no
other bedside monitors with the same Bed ID/OR ID. If there are more
than one bedside monitors with the same OR ID, the duplicated bedside
monitors cannot be monitored on the central monitor.
z
When connecting to the DS-LAN
ΙΙ
network, set the OR ID in the range
from “001” to “048”.
N O T E
When the monitor indicates that the measurement data is out of range
(“
×××
” display), the minimum or maximum value of the range will be
displayed at the central monitor.
【
Out of range
】
【
Central Monitor
】
HR
301bpm or above
300bpm
RR
151Bpm or above
150Bpm
BP –51mmHg
or
below
–50mmHg
301mmhg or above 300mmHg
TEMP –0.1
°
C or below
0
°
C
50.1
°
C or above
50.0
°
C
8
W
ire
d Ne
tw
ork
C
on
str
uc
tion
8
-
19