To Set ON/OFF of Arrhythmia Alarm
Selecting ON will generate arrhythmia alarm.
Selecting OFF will not generate arrhythmia alarm.
Page
Item
Selection
Asystole ON
(ON)
VF
ON (ON, OFF)
VT
ON (ON, OFF)
Slow VT
ON (ON, OFF)
Tachy ON,
OFF
Page 1/2
Brady ON,
OFF
Run ON,
OFF
Couplet ON,
OFF
Bigeminy ON,
OFF
Trigeminy ON,
OFF
Pause ON,
OFF
Page 2/2
Frequent ON,
OFF
●
To Set the Arrhythmia Detection Level
Select the level to detect each arrhythmia.
1
Pressing the detection level key (ex. 5 for Asystole) for each arrhythmia will display the
window to adjust the detection level.
2
Set the detection level.
Item
Range
Asystole
3 to 10 sec.
Run
2 to 8 beats
Pause
1.5 to 5 sec.
Frequent
1 to 50 beats/min.
3
Close the window to set the detection level.
Press the Close key.