W A R N I N G
The patient classification affects the measurement accuracy of QRS
detection, NIBP measurement, etc. Make sure to select the correct
classification for the monitoring patient.
1
Select the patient classification from Adult / Child / Neonate .
This selection is synchronized with the patient classification
selection on the patient admit/discharge menu.
If the patient classification is changed here, the selection on
the admit/discharge menu will be automatically changed to
the same selection.
2
The selected patient classification will be displayed on the Home Display.
●
Quick Measurement
1
Select ON/OFF for “Quick Measurement”.
ON will perform the measurement for duration of about 20
to 25 seconds in case of an adult patient.
N O T E
The quick measurement setting will be effective when the patient classification
is adult or child. If the patient classification is neonate,
normal
NIBP
measurement will be performed regardless of the quick measurement setting.