O Compensation
If N
2
O is present in the respiration circuit, the CO
2
value tends to be displayed higher than the actual
value. By setting the N
2
O compensation ON, this can be adjusted.
Select ON or OFF for N
2
O compensation.
●
Anesthetic Gas Compensation
By inputting the anesthetic gas concentration, compensation can be made to display more accurate
value.
Pressing the key will display the tool box to set the
value.
Use the
Ï
Ð
keys to adjust the anesthetic gas
compensation (anesthetic gas concentration).
Press the Close key after setting the anesthetic gas
compensation.
N O T E
The value cannot be changed if the total value of O
2
compensation and anesthetic
gas compensation exceeds 100%. In such case, change the anesthetic gas
compensation value after changing the O
2
compensation value.
●
Atmospheric Pressure Compensation
The atmospheric pressure can be adjusted to compensate for pressure differences.
Pressing the key will display the tool box to set the
value.