6
-
46
●
SpO
2
Averaging
The averaging time for SpO
2
value can be selected.
W A R N I N G
Be cautious when setting the "SpO
2
Averaging" duration as the SpO
2
alarm is
based on the displayed SpO
2
value which is averaged from the duration set in
"SpO
2
Averaging". The alarm occurrence time will be affected or may not
occur for the transient value of SpO
2
depending on the set duration.
1
Press the Setup key.
The setup menu to select the SpO
2
averaging time will be
displayed.
2
Select the averaging time from 2-4sec. / 4-6sec. / 8sec. / 10sec. / 12sec.
/ 14sec. / 16 sec. .
●
Synchronization Mark/Tone
A mark synchronized to HR or PR can be displayed inside the numeric data box.
A synchronization tone will be also generated.
1
Select HR , PR or OFF for “Sync. Mark/Tone”.
HR will display a mark synchronized to HR and will
generate HR synchronized tone.
PR will display a mark synchronized to BP or SpO
2
according to the priority set for the PR source. PR
synchronized tone will be also generated.
OFF will not display a synchronized mark and will not
generate a synchronized tone.
Sync. will display a mark and generate a tone synchronized
to HR, BP, or SpO
2
according to the alarm source.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...