7
-
68
Other Monitor Data
Display/Alarm
This section explains about the function to display the waveform and numeric data and to set alarms for
other monitors.
To use this function, a DS-LAN
ΙΙ
wired network connection is required.
Other Monitor Display
The other monitor display can be accessed from the menu or from the preprogrammed user key.
Also, by setting the other monitor alarm ON, Other Mon. key will be displayed when other monitor
generates an alarm. By pressing this Other Mon. key, the display for the other monitor can be
accessed.
1
Press the Menu
→
Function
→
Other Monitor keys to display the other monitor
selection menu.
On the other monitor selection menu, select the Operation Room
ID to display from the beds connected to the wired network.
For the DS-LANII network, there are 48 beds selection, and for
the DS-LANIII network, there are 100 beds selection.
The Operation Room ID for the alarm generating monitor will be
displayed in red.
The monitor displaying this menu will be displayed in gray.
The key LED for the selected monitor will be lighted.
Display from the
Other Mon. key.
Display from the menu
Display from the user key