Maintenance Check
Daily and Periodic Check
This section explains daily check and periodic check items for the device.
About Maintenance Check
Periodic inspection must be performed. When reusing the device which was left unused for a while,
always check that the device operates properly and safely before use.
To ensure safety, reliability, and high performance, a “Daily Check” and “Periodic Inspection” must be
performed. We are not liable for any accident arising from lack of maintenance.
C A U T I O N
z
Do not open the housing of this device.
z
Avoid alcohol or other liquids from getting into the device.
●
Daily Check
Perform daily inspection using the “Daily Check List” on the next page.
●
Periodic Check
The safety check conformed to EN 60601-1 must be performed at least once every 2 years.
Periodic inspection of medical electronic equipment is mandatory to prevent failures and accidents and
to ensure safety and reliability.
Periodic maintenance may be performed by each medical institution or by a third party by concluding a
“Maintenance Contract”.
For more details, contact your nearest service representative.
Periodic Replacement Parts
10
Ma
in
te
na
nc
e C
heck
To ensure reliability of safety, function, and performance for this device, the following components
must be replaced periodically. When replacing, contact our service representative.
Short Term Backup Battery
Periodic Replacement Period: 3 years
NIBP Unit
Periodic Replacement Period: 100,000 times of measurement
Recorder Unit
Periodic Replacement Period: 350 hours (Rec. accumulated time)
CO
2
Absorber (MGU-701/702/Poet IQ)
Periodic Replacement Period: 1year
O
2
Cell (MGU-701/Poet IQ)
Periodic Replacement Period: 1year
IRMA O
2
Sensor
Periodic Replacement Period: 4 months (even if not used.)
DRYLINE™ Water Trap (MGU-801P/802/803)
Periodic Replacement Period: 1 month
DRYLINE™ Receptacle (MGU-801P/802/803)
Periodic Replacement Period: 1year
N O T E
The display panel utilizes exclusive fluorescent light for its backlight.
As this fluorescent light tube has a product life cycle, it needs to be replaced
periodically. If the display becomes dark, flickers, or does not light, contact your
nearest service representative.
C A U T I O N
The service life of the cooling fan used for this equipment is 40,000 hours.
However, if it is used with extremely high frequency for long duration, a
message such as “FAN Error” or “DS-7000 Hardware Error” may be displayed
or unusual sound may be generated from the cooling fan. In such case, contact
your nearest service representative.
10
-
9
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...