Telemetry
There is no reception at the Central Monitor.
Cause
: The channel ID or group ID do not correspond with the telemetry receiver.
Solution : Set the correct channel ID and group ID on the DS-7000.
The impedance respiration waveform is not shown at the Cental Monitor.
Cause 1 : On the respiration setup, “RR Alarm Source” is set to GAS
.
Cause 2 : On the respiration setup, “RR Alarm Source” is set to Ventilator
.
Solution : Set the “RR Alarm Source” to Impedance .
The BP waveform of 100mmHg or above cannot be properly transmitted.
Cause
: The BP waveform and scale do not correspond.
Solution : When the BP waveform is above 100mmHg, set the BP scale above 100mmHg on the
DS-7000.
Remote Control Unit
The remote control unit does not operate.
Cause 1 : The Remote Control Monitor ID does not correspond with the remote control unit.
Solution : Set the correct Remote Control Bed ID.
Cause 2 : The section does not correspond with the remote control unit.
Solution : Set the correct section.
Proper operation cannot be performed using the remote control unit.
Cause 1 : The user key on the remote control unit and remote control setup on the monitor do not
correspond.
10
Solution : Set the remote control setup on the monitor so that it corresponds to the remote control
user key.
Trou
ble
sho
oting
General
Nothing is displayed but the main power indicator is lighted in red.
Cause
: A system error has occurred.
Solution : Turn off the power, unplug the power cable, and contact our service representative.
The “Adjusting” message is displayed. Numbers are displayed large on the screen.
Cause
: This is the test mode. Stop using the device immediately.
Solution : Restart the system. The test mode will be cancelled.
If the same situation is observed again, contact our service representative and turn off the
DIP switch No. 6.
The data is initialized each time the power is turned ON.
Cause 1 : The internal switch is set to initialize.
Solution : The internal switch setting needs to be changed. Contact our service representative and
set the rotary switch to 0.
Cause 2 : The battery for backup memory is depleted.
Solution : The battery needs to be replaced. Contact our service representative.
The display is not clear.
Cause 1 : The display brightness is not adjusted.
Solution : Due to the LCD display characteristic, the visible range is limited. Adjust to the
appropriate brightness.
10
-
25
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...