Pulse Wave Size
1
Select the waveform size.
Select the waveform size for displaying and recording from
×
1/4 ,
×
1/2 ,
×
1 ,
×
2 ,
×
4 .
SpO
2
Alarm
1
Press the SpO
2
Alarm key to display the alarm setup menu.
Select ON/OFF for SpO
2
alarm, and set the upper and lower alarm limit.
Also, when the SpO
2
value is unstable around the lower alarm limit, the frequently generated alarm
can be corrected by setting the SEC (second) alarm function.
The upper and lower limits can be set in 1% increment.
Key
Item
Description
Individual Alarm
Selecting ON will generate the SpO
2
alarm.
Selecting OFF will not generate the SpO
2
alarm.
▲
Current Value
Indicates the current value.
Lower Alarm Limit
Sets the lower alarm limit (70 to 99%).
Setting the value to 70% or below will turn OFF the alarm.
Upper Alarm Limit
Sets the upper alarm limit (71 to 100%).
Setting the value to 100% or above will turn OFF the
alarm.
Automatic Setup
Automatically sets the upper limit to OFF, and the lower
limit to 90%.
N O T E
Whether to use the SEC (second) alarm function and its threshold selection should
be based on the patient’s clinical indication portent and medical evaluation.
To maintain the alarm setting even after the power is turned OFF or after the discharge procedure,
store the setting to one of the alarm modes, or select ON for “Backup at Discharge” on the
Discharge Mode setup menu under the Initial Settings.
Reference
For the alarm mode setup procedure, refer to “9. Initial Settings –Alarm Mode Setup–”.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...