6
-
18
ON/OFF of Parameter Display
1
Press the Display ON/OFF key.
The confirmation display for the ON/OFF of ECG display will
appear.
2
Select Display ON or Display OFF .
Pressing the Display ON key will display the waveform and
numeric data.
Pressing the Display OFF key will not display the
waveform and numeric data.
When ECG electrodes are attached to the patient with the ECG display set to OFF, the ECG
waveform and numeric data will be automatically displayed after 10 seconds.
C A U T I O N
When the waveform and numeric data display is set to OFF, the alarm
generation and table/ trend data input will also be suspended.
The Disp. OFF message will be displayed
inside the numeric data box.