Automatic Lead Switching
When the Lead-Off condition occurs, the “Check Electrodes” message will be displayed and a new
ECG lead will be automatically set if the “Automatic Lead Switching” is set to ON.
The automatic lead switching will be performed for ECG 1 and ECG 2.
Lead Switching when Electrode is Detached
Auto Lead Selected
Lead Type
Detached
Electrode
ECG1 ECG2
R
ΙΙΙ
ΙΙΙ
4-Electrode
Lead
L
ΙΙ
ΙΙ
R/R+C
ΙΙΙ
ΙΙΙ
L/L+C
ΙΙ
ΙΙ
5-Electrode
Lead
C
ΙΙ
aVR
1
Press the Page Down key to display the 2nd page of the ECG setup menu.
Select ON or OFF for “Auto Lead Switch”.
ON will automatically switch to another lead when an
electrode is detached.
OFF will not switch lead when an electrode is detached.
●
Synchronization Mark/Tone
A mark synchronized to HR or PR can be displayed inside the numeric data box.
A synchronization tone will be also generated.
HR Mark
1
Select HR , PR or OFF for “Sync. Mark/Tone” on the 2nd page of the ECG setup menu.
HR will display a mark synchronized to HR and will
generate HR synchronized tone.
PR will display a mark synchronized to BP or SpO
2
according to the priority set for the PR source on the BP setup
menu or SpO
2
setup menu. A PR synchronized tone will be
also generated.
OFF will not display a synchronized mark and will not
generate a synchronized tone.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...