4
-
57
●
To Set the HR Low Limit for VT
Set the analysis
condition to detect VT.
Select the HR low limit to detect VT from 120 or 140bpm.
If HR is below the set value, it will be detected as
Slow_VT.
●
To Set the HR Low Limit for RUN
Set the analysis
condition to detect RUN.
1. Pressing the detection level key (ex. 40 for the following screen) will display the window to
adjust the detection level.
2. Set the detection level.
Use the arrow keys to set the detection threshold.
It can be set in the range from 0 to 100bpm.
3. Close the window to adjust the detection level.
Press the Close key.
C A U T I O N
z
The settings for the “HR Low Limit for VT” and “HR Low Limit for RUN” will
be compared with the average HR of continuous VPC. Therefore, the
displayed HR value at alarm generation may be lower than the settings if it
is just after the VT detection, or if RUN with few continuous VPC is
detected.
z
When a wired network (DS-LANII/DS-LANIII) is constructed, the setups for
“HR Low Limit for VT” and “HR Low Limit for Run” cannot be performed on
some central monitors depending on the model type or software version.
4
Alarm Setup