3
Select the recall factor (arrhythmia).
Pressing the Arrhythmia key will display the menu to select the arrhythmia
alarm factor.
Select the arrhythmias for recall factors. The keys with the LED
lighted will be the recall factors.
The alarm OFF mark will be displayed inside the arrhythmia
key if the alarm is set to OFF for that arrhythmia.
z
The recall waveform will start with the following delay time tracing back from
the alarm occurrence.
Neonate
Adult
Child
Meas. Data
Alarm
Arrhy. Alarm
Delay Time
12 sec.
12 sec.
8 sec.
12sec.
N O T E
z
For multigas data alarm, the delay time is 8 seconds.
7
Recall