6
-
73
MAC Value Calculation Constant
The calculation constant for MAC value can be set.
The MAC value is calculated from the following formula.
Ex N
2
O Ex PAGT Ex SAGT
MAC =
+
+
X(N
2
O) X(PAGT) X(SAGT)
Ex N
2
O : Exp N
2
O (%)
Ex PAGT : Exp Primary Agent (%)
Ex SAGT : Exp Secondary Agent (%)
X(N
2
O) :
N
2
O Constant
X(PAGT) : Primary Agent Constant
X(AGT) : Secondary Agent Constant
N O T E
The MAC value will not be displayed unless ON is selected for “MAC Display”.
Perform the setup as necessary.
1
Press the Menu
→
Initial Settings
→
MAC Value keys.
The MAC value setup menu will be displayed.
If you want to change the displayed default value, input the
numbers using the numeric keypad.
Then, press the key for the corresponding constant.
6
Multiga
s Data
(MGU-7
01/7
02/Poet IQ, MGU-80
1P/8
02/803
)