11
-
9
For Gas Mixtures of 2% Sevoflurane
Agents
Agent Volume
Change of CO
2
N
2
O 60%
0.0
Halothane 4%
+0.3
Enflurane 5%
+0.6
Isoflurane 5%
+0.2
Sevoflurane 5% N/A
Desflurane* 5% –0.2
Helium 50%
0.0
Ethanol 1%
+0.1
Isopropanol 1% +0.1
Acetone 1%
+0.1
Methane 1%
0.0
Metered dose inhaler propellant
1%
+0.1
80% Xenon
Not intended for use with Xenon
N/A
For Gas Mixtures of 2% Desflurane
Agents
Agent Volume
Change of CO
2
N
2
O 60%
0.0
Halothane 4%
+0.5
Enflurane 5%
+0.4
Isoflurane 5%
+0.5
Sevoflurane 5% –0.1
Desflurane* 5% N/A
Helium 50%
–0.1
Ethanol 1%
0.0
Isopropanol 1% 0.0
Acetone 1%
0.0
Methane 1%
–0.1
Metered dose inhaler propellant
1%
+0.2
80% Xenon
Not intended for use with Xenon
N/A
For Gas Mixtures of 40% N
2
O
Agents
Agent Volume
Change of CO
2
N
2
O 60%
N/A
Halothane 4%
+1.0
Enflurane 5%
+1.2
Isoflurane 5%
+1.3
Sevoflurane 5% +1.0
Desflurane* 5% +1.2
Helium 50%
+0.4
Ethanol 1%
+0.8
Isopropanol 1% +1.5
Acetone 1%
–0.1
Methane 1%
+0.5
Metered dose inhaler propellant
1%
+0.3
80% Xenon
Not intended for use with Xenon
N/A
*
Presence of 15% Desflurane in a mixture with other halogenated gases will result in the display of
15% Desflurane readings or greater.
Other Interference
Ether, cyclopropane and methoxyflurane are contraindicated for use with the multigas unit. The effect
of ethyl alcohol and metabolic ketones and acetone are negligible.
W A R N I N G
The multigas unit is not intended for monitoring gas mixtures containing
methoxyflurane or halogenated hydrocarbons not specifically listed as a
monitored gas.
11
Specifi
cation /
Performan
ce
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...