6
-
29
BP Scale
1
Select the scale on the BP setup menu.
Select the full scale for displaying and recording.
The scale selection will differ depending on the label.
BP Label
Scale
20, 50, 75, 100, 150, 200, 250, 300 (mmHg)
BP1 to BP6, UAP,
User Label
4, 8, 12, 16, 20, 24, 32, 40 (kPa)
50, 75, 100, 150, 200, 250, 300 (mmHg)
ART, LVP, IAP
8, 12, 16, 20, 24, 32, 40 (kPa)
20, 40, 50, 75, 100, 150 (mmHg)
PAP, RVP
4, 6, 8, 12, 16, 20 (kPa)
10, 20, 30, 40, 50, 100 (mmHg)
2, 4, 5, 6, 8, 16 (kPa)
CVP, RAP, LAP
20, 40 (cmH
2
O)
* only for CVP
5, 10, 15, 20, 50, 100 (mmHg)
ICP
1, 2, 3, 4, 8, 16 (kPa)
C A U T I O N
The BP waveform with a scale above the programmed scale will not be
properly transmitted on a wireless network. Select a proper scale for the
waveform.
6
Inva
sive Bloo
d Pressu
re