NIBP (Non-Invasive Blood Pressure)
(IEC 60601-2-30: 1999 Medical electrical equipment- Part 2: Particular requirements for the
safety of automatic cycling indirect blood pressure monitoring equipment)
(EN 60601-2-30: 2000 Medical electrical equipment- Part 2: Particular requirements for the
safety of automatic cycling indirect blood pressure monitoring equipment)
(AAMI SP10: 2002+A1:2003+A2:2006 Manual, electronic or automated sphygmomanometers)
Method :
Oscillometric
Method
Measurement Range
: Adult
10 to 280 mmHg/1.3 to 37.3kPa
Child
10 to 180mmHg/1.3 to 24.0kPa
Neonate 10 to 130mmHg/1.3 to 17.3kPa
Resolution :
1mmHg
Static Pressure Accuracy :
±
3 mmHg/0.4kPa
Accuracy
Maximum Standard Deviation : 8mmHg/1.0kPa
Maximum Average Error
:
±
5mmHg/0.6kPa
PR Measurement Range : 40 to 240bpm
±
2% or
±
2bpm (whichever greater)
Deflation Speed
: 5
±
1mmHg/sec (Quick Measurement OFF)
10
±
2mmHg/ sec (Quick Measurement ON)
Safety Mechanism
: Adult
300mmHg or above
Child 210mmHg
or
above
Neonate 150mmHg
or
above
Gas Concentration (Multigas Unit: MGU-701/MGU-702, optional)
The performance is according to the Criticare Systems Inc. OEM HIGH IQ
TM
AGENT MODULE
specification.
Method
: Sidestream; Non-dispersive infrared (NDIR)
Identified Gases
: CO
2
, N
2
O, O
2
, Halothane, Enflurane, Isoflurane, Desflurane, and
Sevoflurane
Flow Rates
: 100mL/min, 150mL/min, 200mL/min
Measurement Range : CO
2
0 to 12.5% (0 to 99mmHg, 0 to 12kPa)
N
2
O
0 to 99%
O
2
0 to 100%
Halothane:
0 to 10%
Enflurane:
0 to 10%
Isoflurane:
0 to 10%
Sevoflurane:
0 to 10%
Desflurane:
0 to 20%
RR
1 to 60 Bpm
Accuracy :
CO
2
:
±
0.2% abs. or 4% of reading
N
2
O:
±
(1.5% abs. +4% of reading)
O
2
:
±
3 vol. % (0 to 90%)
±
4 vol. % (91 to 100%)
Agents:
±
(0.1% abs. +4% of reading)
RR:
±
2 Bpm or 2% of reading
Time to Detect Agent <15 second@200mL/min
Agent Detection
Resolution
: 0.1 Volume Percent
Detection Threshold
: Halothane 0.2 vol. %
Enflurane
0.3 vol. %
Isoflurane
0.3 vol. %
Sevoflurane 0.3 vol. %
Desflurane 0.3 vol. %
Mixed Gas Threshold : 0.2 vol. % + 10 % of total concentration
Rise Time
: CO
2
: 350msec
N
2
O: 400msec
O
2
:
600msec
Agents: 450msec
Occlusion Clearing
: Automatic
Auto Zeroing
: Occurs 30 to 60 minutes
Duration: 3.0 to 7.0 seconds
Manual user calibration not required
Warm-up Time
: 1 minute to first waveforms
<20 minutes to full accuracy
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...