ST Alarm Setup
The ST upper value and lower value compared with the reference waveform will be set.
The alarm value is to be set for each measurement unit (mm / mV). The upper and lower limit can be
set in 1mm / 0.1mV increment.
1
Press the Menu
→
Function
→
ST
→
Alarm keys to display the alarm setup menu.
2
Select ON/OFF for “ST All Alarm”.
ON will generate a ST alarm.
However, the alarm will not be generated for the lead
where the individual alarm is set to OFF .
OFF will not generate a ST alarm.
3
Select the lead to set the alarm limit.
Press one of the lead keys to set the alarm limit.
4
Set the upper and lower alarm limit.
Use the
Ï
,
Ð
keys to adjust the alarm limit.
Item
Description
Lower Alarm Limit
Select a lower alarm limit (–20.0 to 19.0mm/–2.00 to 1.90mV).
Alarm will be set to OFF if the value –20mm/–2.0mV or lower is
selected.
Upper Alarm Limit
Select an upper alarm limit (–19.0 to 20.0mm/–1.90 to 2.00mV).
Alarm will be set to OFF if the value +20mm/+2.0mV or above is
selected.
5
Select ON/OFF of “Indiv. Alarm”.
ON will generate a ST alarm for the selected lead.
OFF will not generate a ST alarm for the selected lead.
6
Select “Auto” for automatically setting the alarm limit.
Pressing the Auto key will automatically set the upper alarm limit to current
ST value +0.2mV (+2mm), and lower alarm limit to current ST value –0.2mV
(–2mm).
Selecting “Auto” will automatically turn ON the ST alarm.
The limits will be automatically set even if the upper or lower limit is set to OFF.
Upper limit cursor
Lower limit cursor
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...