Concentration
(HPD-800: Capnostat5)
The GAS setup menu can be accessed by pressing the GAS Setup key.
●
ON/OFF of Gas Measurement
Select ON or OFF for “Gas Measurement”.
ON will perform gas measurement.
OFF will not perform gas measurement.
●
EtCO
2
Peak Picking Duration
The duration to pick the EtCO
2
maximum value can be selected from 10 sec., 20 sec., or OFF.
10sec , 20sec will display the maximum EtCO
2
value for the selected duration.
OFF will display the EtCO
2
value for each respiration.
As the EtCO
2
value display is updated each second,
EtCO
2
value for each respiration cannot be displayed if
respiration rate is 60Bpm or above.
●
O
2
Compensation
By inputting the O
2
concentration value, compensation can be made to display more accurate value.
Pressing the key will display the tool box to set the
value.
Use the
Ï
Ð
keys to adjust the O
2
compensation
(O
2
concentration).
The O
2
compensation can be set in 1% increment for
the value up to 30%, and 5% increment for the value
30% and above.
Press the Close key after setting the O
2
compensation.
N O T E
The value cannot be changed if the total value of O
2
compensation and anesthetic
gas compensation exceeds 100%. In such case, change the O
2
compensation value
after changing the anesthetic gas compensation value.
6
CO
2
Co
ncent
ration (HP
D
-8
00: Cap
no
sta
t5)
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...