4
-
27
PCWP Value, PCWP Measured Time
When the BP label is PAP, PCWP (Pulmonary Capillary Wedge
Pressure) and the measured time can be displayed.
PDP Value
When the BP label is IAP, PDP (Peak Diastolic Pressure) of IABP can
be displayed.
Systolic Pressure (SYS) = Peak Systolic Pressure (PSP).
CPP Value
When the BP label is ICP, and artery pressure is labeled as ART,
CPP
(
Cerebral Perfusion Pressure
)
can be measured. CPP = Mean
Value of Arterial Pressure – Mean Value of Intracranial Pressure
If the CPP value is less than 0, ICP or ART has not been measured,
or the zero balance has not been performed for ICP or ART, the value
will be displayed as “
−
−
−
”. Also, the alarm cannot be set for CPP.
Mean Wave
Displayed when mean waveform is set to ON (BP configuration
menu.).
TEMP Label
The label set for the temperature will be displayed.
TEMP Value
Displays the temperature measurement value. The 400 series
temperature probe can be used. When the measurable range is
exceeded, “xxx” will be displayed.
Blood Temperature
By using the thermodilution catheter for the CO measurement, blood
temperature can be displayed. When the measurable range is
exceeded, “xxx” will be displayed.
Gas Unit Data
When a gas unit is connected, the gas data (CO
2
/ anesthetic gas / O
2
/ N
2
O concentration) will be displayed.
When mixed anesthetic gas is used, primary anesthetic gas
concentration, secondary anesthetic gas concentration will be
displayed.
When “MAC Value Display” is set to ON, the MAC value will be
displayed.
The gas unit data can be individually displayed inside each numeric
data box.
・
InspCO
2
value / EtCO
2
value
・
FiO
2
value / ExO
2
value
・
InspN
2
O value / ExN
2
O value
・
InspAGT value / ExAGT value
The CO
2
measurement unit can be selected from mmHg/kPa/% on
the “Initial Settings” menu.
Ventilator Measurement
When a ventilator is connected, the ventilator measurement data will
be displayed.
4
De
scription of
the Displ
ay
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...