HR/PR Alarm Source
1
Set the “HR/PR Alarm Source” on the 2nd page of the ECG setup menu.
HR will generate the alarm based on HR.
PR will generate the alarm based on BP or SpO
2
according
to the priority set for the PR source on the BP setup menu or
SpO
2
setup menu.
Auto will automatically select the alarm source according to
the priority set on the “Source Setup” (Initial Settings).
W A R N I N G
The HR/PR alarm will not be generated unless the numeric data box
corresponded to the selected HR/PR alarm source is displayed. When the
HR/PR alarm source selection is changed, be sure to display the numeric
data box corresponded to the selected HR/PR source.
C A U T I O N
If “HR/PR Alarm Source” is PR, and “PR Source” is BP, the ECG waveform
will not be transmitted on a wired network. On the central monitor, PR-IBP
value will be displayed for HR. However, the HR value from ECG will be
displayed on the NIBP list and ST measurement table.
N O T E
z
The synchronized mark will be displayed inside the numeric data box for
the selected “Sync. Mark/Tone” source regardless of the “HR/PR Alarm
Source” setting.
z
The “Sync. Mark/Tone” and “HR/PR Alarm Source” settings will be
synchronized with the HR/PR user key function.
●
ST/VPC/Arrhy. Alarm Display
1
Set the “ST/VPC/Arrhy. Alarm Display” on the 2nd page of the ECG setup menu.
ON will display the ST, VPC, arrhythmia alarm factor on the
ECG numeric data box.
OFF will not display the ST, VPC, arrhythmia alarm factor
on the ECG numeric data box.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...