Cleaning the Housing
Clean the housing using tightly squeezed gauze or an absorbent cotton cloth dampened with alcohol
or a neutral cleanser.
C A U T I O N
z
Clean the equipment frequently so stains can be removed easily.
z
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
z
Do not allow liquids or cleaning solution to enter the monitor or connectors.
z
Do not use organic solvents, thinner, toluene or benzene to avoid damaging
the resin case.
z
Do not polish the housing with abrasive or chemical cleaner.
z
When sterilizing the entire room using a spray solution, pay close attention
not to have liquids get into the monitor or connectors.
z
Use only neutral detergent to clean the housing. Do not use a chemical
cloth, scrub brush, abrasive, polishing powder, hot water, volatile solvent or
chemicals (cleanser, thinner, benzine, benzol, and synthetic detergent for
house and furniture), or sharp-edged tools. The surface resin coating may
be damaged, resulting in discoloration, scratches, or other problems.
Cleaning the NIBP Cuff and Air Hose
Take out the bladder from the cuff cloth bag, and wash the cloth bag with neutral detergent. After
drying it, put the bladder back in.
Disinfecting the Blood Pressure Transducers
Disinfect the blood pressure transducers according to the manufacturer’s guidelines.
10
Cleaning and Disinfecting the SpO
2
Sensor
Clea
ning
【
NELLCOR
®
Sensor
】
y
Do not soak the sensor in water or antiseptic solution.
y
Wipe the DURASENSOR with disinfectant such as 70% alcohol. Do not disinfect by applying
radioactive rays, steam, or ethylene oxide.
y
OXISENSOR is a disposable sensor. Do not reuse or resterilize.
【
MASIMO
®
Sensor
】
y
Do not immerse the sensor or patient cable in water or cleaning solution. (Sensor and connector are
not waterproof.)
y
Do not disinfect the sensor and cable by irradiation, steam, or ethylene oxide.
y
Clean the Masimo reusable sensor (LNOP
®
DCI) and patient cable using the following procedure.
(1) Remove the sensor from the patient. Disconnect the patient cable from the sensor.
(2) Disconnect the sensor from the DS-7000M.
(3) Wipe the sensor and cable using 70% isopropyl alcohol cotton.
(4) Before reusing, dry it completely with air.
y
The Masimo disposable sensor can be reused on the same patient if the light emitting and receiving
part is clean, and if it is still adhesive to the skin.
The adhesiveness will return by completely drying the sensor after cleaning with alcohol.
y
Cleaning, Disinfecting, and Sterilizing the Temperature Probe
y
Clean/Disinfect/Sterilize the temperature probe according to the guidelines provided with the probe
product.
10
-
5
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...