Precautions about the Maintenance
Safety Inspection and Maintenance
For safe operation of the equipment, regular inspection and maintenance is required. Once a year,
check all cables, devices, and accessories for damage, earth impedance, earth and leakage currents,
and all alarm functions. Also, ensure that all safety labels are legible. Maintain a record of these safety
inspections.
Immediate maintenance has to be carried out if ;
y
the equipment was subjected to extreme mechanical stress, e.g. after a heavy fall.
y
the equipment was subjected to liquid spill.
y
the monitoring function is interrupted or disturbed.
y
parts of the equipment enclosure are cracked, removed, or lost.
y
any connector or cable shows signs of deterioration.
Reference
Refer to “10. Maintenance” for details.
W A R N I N G
Never open the housing while the equipment is in operation or connected to
hospital grade outlet as it may result in electric shock.
Maintenance, Modifications, and Repairs
Fukuda Denshi is liable for the safety, reliability, and performance of its equipment only if;
y
Maintenance, modifications, and repairs are carried out by authorized personnel.
y
Components are used in accordance with Fukuda Denshi operating instructions.
A full technical description of the DS-7000 is available from your local Fukuda Denshi representative.
xii
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...