DRYLINE™ WaterTrap:
Emptying interval (half full, worst case),
Adult
: 17 h @ 200 ml/min, 37
°
C, 100% RH
Neonate : 20 h @ 120 ml/min, 37
°
C, 100% RH
Gas Concentration (HPD-800 with RESPIRONICS
®
Capnostat 5, optional)
Method
: Infra-Red Solid-State Method, Mainstream Method
Measurement Range
: 0 to 150mmHg
Initialization Time
: Within 15 seconds until the measurement can be performed
Within 2 minutes until maximum accuracy is achieved
Measurement Accuracy
: 0 to 40mmHg: ± 2mmHg
41 to 70mmHg: ± 5%
71 to 100mmHg: ± 8%
101 to 150mmHg: ± 10%
Response Time
: Airway Adapter (Adult) 60ms and below
Airway Adapter (Pediatric) 50ms and below
RR Measurement Range
: 0 to 150Bpm
RR Measurement Accuracy : ± 1Bpm
Gas Concentration (PhaseIn IRMA, optional)
According to PhaseIn IRMA specification.
Method
: Infra-Red Absorption Method (Excluding O
2
)
Galbanic Oxygen Sensor
Parameter :
End-Tidal
CO
2
concentration (EtCO
2
) / Inspiratory CO
2
concentration
(InspCO
2
) / Respiration Rate / N
2
O concentration, O
2
concentration,
Anesthetic Agent (AG) Concentration
Warm-up Time
: Less than 10 sec. until measurement can be performed.
1 min. until maximum accuracy is achieved.
Fi and ET Values : IRMA CO
2
: CO
2
IRMA ICU: CO
2
, O
2
IRMA OR: CO
2
, N
2
O, O
2
, primary agent (HAL, ISO, ENF, SEV, DES)
IRMA AX: CO
2
, N
2
O, primary agent (HAL, ISO, ENF, SEV, DES)
IRMA AX+:CO
2
, N
2
O, primary agent (HAL, ISO, ENF, SEV, DES),
secondary agent (HAL, ISO, ENF, SEV, DES), agent ID
Measurement Range and Accuracy (After warm-up time when maximum accuracy is achieved):
Measurement Accuracy
Parameter
Measurement
Range
*1
Standard Condition
Under Influence of
Other Gas
RR
0, 3 to 150Bpm
–
–
0 to 10%
±
(0.2%
ABS
+2%
REL
)
±
(0.3%
ABS
+4%
REL
)
CO
2
10 to 20%
Unspecified
Unspecified
0 to 10%
Unspecified
Unspecified
O
2
10 to 100%
±
(1%
ABS
+2%
REL
)
±
(2%
ABS
+2%
REL
)
N
2
O
0 to 100%
±
(2%
ABS
+2%
REL
)
±
(2%
ABS
+5%
REL
)
0 to 5%
±
(0.15%
ABS
+5%
REL
)
HAL, ISO, ENF
5 to 12%
Unspecified
0 to 8%
±
(0.15%
ABS
+5%
REL
)
SEV
8 to 15%
Unspecified
0 to 18%
±
(0.15%
ABS
+7%
REL
)
DES
18 to 22%
Unspecified
Single Agent
*2
±
(0.2%
ABS
+10%
REL
)
Dual Agent
*3
±
(0.2%
ABS
+10% of total
agent concentration)
Primary agent
threshold
0.15vol%
Secondary agent
threshold
0.2vol%+10% or total
agent concentration
*
1 : Gas concentration reported in units of volume percent.
*
2 : The accuracy specification is valid for all specified environment conditions, and includes
cross effects from the following interfering gases: N
2
O, O
2
, Hal, Iso, Enf, Sev, Des,
Ethanol, Isopropyl alcohol, Acetone, Methane, He.
*
3 : The accuracy specification for IRMA AX/OR is not valid if other anesthetic agents than
the selected agent are present in the gas mixture.The accuracy specification for IRMA
AX+ is not valid if more than two agents are present in the gas mixture.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...