ECG Monitoring Condition Setup
Pressing the Setup key on the ECG configuration menu will allow to set the ECG measurement
condition.
●
Filter Mode Selection
The waveform frequency characteristic can be selected from Monitor Mode, ESIS Mode, or Diagnosis
Mode according to the monitoring purpose. Each mode has different frequency characteristic.
The selected filter mode will be printed when recording.
1
Select the filter mode from Monitor , ESIS , or Diag. .
Monitor Mode (Frequency Characteristic: Adult/Child 0.5 to 40Hz, Neonate 1.6 to 40Hz)
This is the standard mode for ECG monitoring. The upper frequency is set to 40Hz to reduce artifact
caused by EMG, etc.
ESIS Mode (When a defibrillation and electrosurgery-proof ECG relay cable is used)
(Frequency Characteristic: Adult/Child/Neonate 1.6 to 15Hz)
The upper frequency is set to 15Hz, so that it can largely reduce the high-frequency artifact.
Diagnosis Mode (Frequency Characteristic: Adult/Child 0.05 to 100Hz, Neonate 1.6 to 100Hz)
Select this mode if ST measurement or high frequency ECG monitoring is performed.
As the lower frequency is set to 0.05Hz, ST level can be accurately measured.
C A U T I O N
The ESIS mode can largely reduce the artifact such as electrosurgery noise
and EMG, but it may also reduce the QRS amplitude. The ESIS mode should
be selected only during electrosurgery.
6
ECG
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...