HR Alarm
1
Press the HR/PR Alarm key to display the alarm setup menu.
Select ON/OFF for HR/PR alarm, and set the upper and
lower alarm limit.
The common alarm value for HR measured from ECG, PR measured from SpO
2
, PR measured from
BP can be set.
The upper and lower limit can be set in 5 bpm increments.
Key
Item
Description
Individual Alarm
Selecting ON will generate the HR/PR alarm.
Selecting OFF will not generate the HR/PR alarm.
▲
Current Value
Indicates the current value.
Lower Alarm
Limit
Sets the lower alarm limit (20 to 295bpm).
Setting a value 20bpm or below will turn OFF the alarm.
Upper Alarm
Limit
Sets the upper alarm limit (25 to 300bpm).
Setting a value 300bpm or above will turn OFF the
alarm.
Automatic Setup
Automatically sets the upper limit to +40bpm, and the
lower limit to –40bpm to the current value.
To maintain the alarm setting even after the power is turned OFF or after a discharge procedure, store
the setting to one of the alarm modes, or select ON for “Backup at Discharge” on the Discharge
Mode setup menu under the Initial Settings.
Reference
For the alarm mode setup procedure, refer to “9. Initial Settings –Alarm Mode Setup–”.
Arrhythmia Alarm
1
Press the Arrhythmia Alarm key to display the arrhythmia alarm setup menu.
ON/OFF of each arrhythmia alarm and analysis threshold level can be set.
Reference
Refer to “7. Function –Arrhythmia Analysis–”
for details.
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...