4
-
64
2
Select the output recorder.
Built-in will record on the DS-7000 built-in recorder.
Central will record on the 3ch recorder connected to the
central monitor.
8ch will record on the 8ch recorder connected to the
wired network.
The central monitor recorder will be the one with the
smallest central ID. 8ch recorder selection can be
performed on the recorder setup menu.
3
Select the waveform for recording.
Up to 3 waveforms can be selected. If the output recorder is
8ch recorder, up to 8 waveforms can be selected.
The waveform position will be automatically adjusted when
recording.
4
Select the duration for recording.
Select the duration from 24sec. or Cont. .
24sec. will automatically stop the recording after 24
seconds.
If 8ch recorder is selected as output recorder, Page can be
selected. Page recording prints out one page.
5
Select the delay time for recording.
None will start the recording from the time the
(Record Start/Stop) key is pressed.
8sec. , 16sec. will start the recording 8 sec. / 16 sec. prior
to the time the key is pressed.
N O T E
If None is selected for the manual recording delay time, QRS classification will not
be printed. To record the QRS classification, select 8sec. or 16sec. for the
delay time.