In case of Q2 Computer, check if the communication is set as follows.
・
Baud Rate: 9600bps
・
Parity Bit: ODD
・
Stop Bit: 1
・
Data Bit: 7
For procedure to check the Q2 Computer communication setup, refer to the operation
manual for the Q2 Computer.
Cause 7 : The software version of Vigilance does not correspond.
Solution : If the Vigilance without the STAT function is connected, the STAT data will not be
displayed. Check the software version of the Vigilance.
An error is caused between the data of Q2 Computer and bedside monitor.
Cause
: Due to difference such as number of significant digit, an error may be caused between
the displayed data and transmitted data of the Q2 Computer. Also, updating of the
monitor data may be delayed due to transmission delay which causes the difference of
value between the Q2 Computer and the DS-7000.
The CO average value is displayed although not measured.
Cause
: The past CO data is stored in OXIMETRIX3, Q-vue, Q2 Computer.
Solution : Clear the stored CO data in the OXIMETRIX3, Q-vue, Q2 Computer before connecting to
an external output cable.
BIS Monitor
The numeric data is not displayed.
Cause 1 : If SQI value is lower than 15, BIS data and SR data will not be displayed.
Solution : Refer to the BIS operation manual and set the SQI value above 15.
10
Cause 2 : The communication setup of the BIS monitor is incorrect.
Solution : ASCII should be set to communicate with the DS-7000.
Check the communication setup and verify that it is set to ASCII.
Refer to the BIS operation manual for procedures.
Trou
ble
sho
oting
Magnetic Card Reader / Bar Code Reader
The magnetic card reader or bar code reader does not function.
Cause 1 : The conversion cable (CJ-756) is not connected.
Solution : If the magnetic card reader or bar code reader is connected directly to the serial port on
this equipment without the conversion cable, it will not function. Make sure to use the
conversion cable.
Radical-7
The numeric data is not displayed.
Cause 1 : The serial output setting on the Radical-7 is not set to ASCII 2.
Solution : Refer to the operation manual of Radical-7, and set the serial output to ASCII 2.
Cause 2 : The software version of the Radical-7 Handheld or Docking Station is not compatible
with the DS-7000.
Solution : For the compatible software version, refer to our service representative.
10
-
31
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...