Respiration
The “CVA Detected” message is displayed.
Cause
: Heartbeat is interfering and superimposed on the respiration waveform.
Solution : Place the electrode as shown below where the heartbeat will be less likely to interfere.
Green
Yellow
Red
“0” is displayed for respiration rate, or apnea alarm is generated.
Cause
: The respiration waveform amplitude is below the detection level (0.2
Ω
).
Solution 1 : Change the electrode site.
Solution 2 : Increase the waveform size.
The respiration waveform and respiration rate is not displayed.
Cause 1 : The ECG relay cable designed for electrosurgical knife is used.
Solution : The impedance respiration can not be measured if the cable designed for electrosurgical
knife is used. Use the standard ECG relay cable if not using the electrosurgical knife.
Cause 2 : The impedance respiration measurement ceased.
Solution : Turn ON the impedance respiration measurement on the ECG setup menu.
Note
: If a pacemaker with the minute ventilation measuring function is used, turn OFF the
impedance respiration measurement. Otherwise, both the pacemaker and the monitor will
not be able to perform accurate measurement.
The measured data is displayed as “
×××
”.
Cause
: The respiration rate is outside the measurement range.
Solution :
・
Check the electrode application.
・
Replace the electrode, or check the lead cable.
Invasive Blood Pressure
The “BP
¼
Transducer OFF” message is displayed.
Cause
: The BP transducer is not connected.
Solution : Connect the transducer.
The “BP
¼
Zero Required” message is displayed.
Cause
: The BP zero balance has not been performed since the power was turned ON.
Solution : Open the three-way valve on the transducer to air and perform a zero balance.
BP value and waveform are not displayed properly.
Cause
: Blood pressure line has not been zero balanced.
Solution : Open the three-way valve on the transducer to air and perform a zero balance.
The measured data is displayed as “
×××
”.
Cause
: The BP value is outside the measurement range.
Solution : Perform a zero balance again.
10
-
16
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...