4
-
61
<
Graph
>
<
Numeric
>
<
OFF
>
N O T E
z
The alarm limit cannot be displayed inside the numeric data box for the following
parameter.
y
GAS
(CO
2
+AGT+O
2
+N
2
O)
y
GAS
(AGT+O
2
+N
2
O)
y
GAS
(N
2
O+AGT)
z
The alarm limit bar graph can be displayed for HR/PR, BP1 or ART, NIBP, SpO
2
,
T1, RR, and CO
2
. It cannot be displayed for other parameters.
6
Select ON/OFF for “Save Data to Table”.
By selecting ON for “Save Data to Table”, the measurements at alarm occurrence can be saved in
tabular format.
7
Select the alarm silence duration for the equipment status alarm.
Sets the alarm silence duration for the equipment status
alarm.
Reference
For types of equipment status alarm, refer to ”Description of Alarm Message and Alarm Sound
●
Equipment Status Alarm Message” in this chapter.
Linked to alarm silence time will silence the alarm when the
(Alarm Silence) key is pressed
for fixed amount of duration for ”Alarm Silence”.
・
If the alarm cause still remains at completion of silence duration, the alarm sound will generate
again.
・
If the same alarm occurs during the alarm silence duration, the alarm sound will not generate.
・
If the new alarm occurs during the alarm silence duration, the alarm sound for the new alarm
will generate.
Linked to each new occurrence will silence the alarm when the
(Alarm Silence) key is
pressed until the situation changes.
・
The alarm will be silenced as long as the alarm cause remains.
・
If the alarm cause is resolved during the alarm silence duration, the alarm silence state will be
cancelled.
・
If the same alarm generates again during the alarm silence duration, the alarm sound will
generate.
4
Alarm Setup
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...