Suspend Monitoring
Suspend / Resume Monitoring
This menu allows suspending and resuming monitoring when a patient temporarily leaves the bed.
Turning the power OFF will erase the recall and ST measurement data, but with this suspend
monitoring function, data measurement, alarm generation, automatic measurement, and automatic
recording can be suspended without erasing any data and setup condition.
To Suspend Monitoring
1
Press the Monitor Suspend key.
The confirmation menu will be displayed.
If you do not want to suspend monitoring, press the
Cancel key to return to the previous display.
2
Monitoring will be suspended.
Pressing the OK key will return to the home display with the Resume key displayed.
The numeric data display and waveform display on the home display will be suspended and only the
Resume key will be effective.
N O T E
z
The telemetry transmission will cease when monitoring is suspended.
Note that square wave will be displayed on the central monitor indicating the too
far condition of the telemetry.
z
The stopwatch counting will continue even when the monitoring is suspended.
5
-
20