HR Average Method
The averaging method of HR measured from ECG can be selected.
1
Select Instant or Average for “HR Average” on the 2nd page of the ECG setup menu.
Instant will display the HR measured from RR interval for
each heartbeat.
Average will display the average HR measured for 6
seconds for adult and child, and 3 seconds for neonate.
●
Noise Detection
1
Set the “Noise Detection” on the 2nd page of the ECG setup menu.
ON will automatically switch the HR numeric data and HR
synchronized tone to PR when electrical surgery noise or
motion artifact is detected.
If PR is not measured, HR data before the noise detection will
remain displayed.
During this condition, HR synchronized tone will not be
generated.
OFF will not switch HR to PR even when noise is detected.
When ON is selected for “Noise Detection”
If some noises are detected and PR is also measured, HR
data will switch to PR.
The measurement data will be surrounded by a red box to
indicate that an artifact is present.
If PR is not measured, then the HR data before the noise
detection remains displayed (with HOLD showing inside the
parameter box).
The measurement data will be surrounded by a red box to
indicate that an artifact is present.
C A U T I O N
Arrhythmia may be erroneously detected as noise.
6
ECG
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...