6
-
47
●
HR/PR Alarm Source
1
Set the “HR/PR Alarm Source”.
HR will generate an alarm based on HR.
PR will generate an alarm based on BP or SpO
2
according
to the priority set for the PR source.
Auto will automatically select the alarm source in the priority
of HR>PR. The PR source (BP or SpO
2
) priority will be
according to the PR Source setting.
Auto will automatically select the alarm source according to
the priority set on the “Source Setup” (Initial Settings).
W A R N I N G
The HR/PR alarm will not be generated unless the numeric data box
corresponded to the selected HR/PR alarm source is displayed. When HR/PR
alarm source selection is changed, be sure to display the numeric data box
corresponded to the selected HR/PR source.
C A U T I O N
If “HR/PR Alarm Source” is PR, and “PR Source” is BP, the ECG waveform
will not be transmitted on a wired network. On the central monitor, the PR-IBP
value will be displayed for HR. However, the HR value from ECG will be
displayed for the NIBP list and ST measurement table.
N O T E
z
The synchronized mark will be displayed inside the numeric data box for
the selected “Sync. Mark/Tone” source regardless of the “HR/PR Alarm
Source” setting.
z
The “Sync. Mark/Tone” and “HR/PR Alarm Source” settings will be
synchronized with the HR/PR user key function.
●
Pulse Wave Detection Sensitivity
The sensitivity to detect the pulse wave can be selected from high or low.
1
Press the Setup
→
Page Down key.
The setup menu to select the pulse wave sensitivity will
be displayed.
2
Select from High or Low .
For standard use, select Low .
To improve the low perfusion condition, or to perform fast tracking when the SpO
2
value changes
abruptly, select High .
C A U T I O N
If High is selected for pulse wave sensitivity, sensor-detached detection
will become somewhat inaccurate.
6
SpO
2
(Ma
sim
o
®
Model; DS
-70
00M
)
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...