6
-
49
2
Select from ON or OFF .
ON will display the perfusion index.
OFF will not display the perfusion index.
●
Signal IQ Wave Display
Whether or not to display the signal IQ wave can be selected.
The signal IQ wave indicates the signal force and pulse wave
timing.
The vertical bar indicates the signal quality. A low vertical line
indicates a bad signal quality.
N O T E
The signal IQ wave cannot be recorded.
1
Press the Setup
→
Page Down key.
The setup menu to set the signal IQ wave will be
displayed.
2
Select ON (display) or OFF (not display).
●
SpO
2
Low Perfusion Alarm
The low perfusion alarm level can be selected from level 3 or level 4.
C A U T I O N
Be cautious not to miss the low perfusion condition when selecting level 4 for
the low perfusion alarm level as the alarm sound will not be generated.
1
Set the alarm level for the low perfusion alarm.
Level3 will generate a single tone alarm sound and display
an alarm message at low perfusion condition.
Level4 will only display an alarm message and will not
generate an alarm sound.
6
SpO
2
(Ma
sim
o
®
Model; DS
-70
00M
)